PRCT150 coating thickness gauge
Price:
After login
Availability:
in stock
Product description
This compact, handy pocket gauge is designed for non-destructive, fast and precise coating thickness measurement. The principal applications lie in the field of corrosion protection. It is ideal for manufactures and their customers, for offices and specialist advisers, for paint shops and electroplaters, for the chemical, automobile, shipbuilding and aircraft industries and for light and heavy engineering.
Features:
- With different external probes, the gauge can be applied to measuring thickness of non-magnetic coating on magnetic metal substrate, as well as non-conductive coating on non-magnetic metal substrate
- Two measuring modes: single or continuous, changeable
- Two work modes: direct or group, changeable
- High precision mode: multiple measurements and automatically data filtering method to reduce disturbance of measure results
- Temperature compensation: compensate the measurement distortion caused by the drifting of temperature
- Give five statistical values: average, maximum, minimum, measure number, and standard deviation
- Two calibration methods can be applied to the gauge; And the system error of the probe can be corrected with the basic probe calibration method
- Data storage: up to 500 measurements storage
- Set boundary: alarm while measurements out of boundary
- Battery information indicates the rest capacity of the battery
- Beep prompting
- Five statistical data:MEAN,MAX,MIN,NO.,S.DEV
- Auto power off function to conserve battery life
Technical specification:
Measuring range: | 0 ~ 1250 μm |
Measuring accuracy: | 3 %H + 1 μm (H - měřená tloušťka) |
Resolution: | 0.1 μm |
Optional probes: | F1 / N1 / F400 / F/90, N1, CN02 |
2 measuring modes: | continous, single |
5 statistical functions: | average, maximum, minimum, no of measurements, standard deviation |
Memory: | 500 measuring results |
Sample min. diameter: | 1.5 mm (probe F), 3 mm (probe N) |
Min. material thickness: | 0.5 mm (probe F), 0.3 mm (probe N) |
Min. Ø base layer: | 7 mm (probe F), 5 mm (probe N) |
Measuring tollerance field: | YES |
Power supply: | batteries AAA, 3 x 1.5 V |
Způsob vypnutí: | automatic, manual |
Power off: | 155 x 68 x 27 mm |
Weight: | 230 g |
Standardní delivery:
- Probe F
- Callibration foil
- Callibration base
- Transportation case
- Operation manual
Optional accessories:
- Optional probes